EasyManuals Logo
Home>Tektronix>Test Equipment>MSO56

Tektronix MSO56 User Manual

Tektronix MSO56
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #892 background imageLoading...
Page #892 background image
Commands listed in alphabetical order
TRIGger:A:HO
LDoff:TIMe
This command sets or queries the A trigger holdoff time. This command is
equivalent to selecting Mode & Holdoff from the Trig menu, selecting Time, and
then setting
the desired Holdoff Time.
Group
Trigger
Syntax
TRIGger:A:HOLDoff:TIMe <NR3>
TRIGger:A:HOLDoff:TIMe?
Related Commands
TRIGger:A:HOLDoff:BY
Arguments
<NR3> species the holdoff time in seconds. The range is from 0 seconds through
10 seconds.
Examples
TRIGGER:A:HOLDOFF:TIME 10 sets the A trigger holdoff time to 10 s.
TRIGGER:A:HOLDOFF:TIME? might return :T RIGGER:A:HOLDOFFTIME
1.2000E-06
, indicating that the A trigger holdoff time is set to 1.2 µs.
TRIGger:A:LOGICQUALication
This command sets or queries the type of logic qualication to perform.
Group
Trigger
Syntax
TRIGger:A:LOGICQUALificat ion {AND|OR|NAND|NOR}
TRIGger:A:LOGICQUALificat ion?
Arguments
AND species to trigger if a ll conditions are true.
NANd species to trigger if any of the c onditions are false.
NOR species to trigger if all conditions are false.
OR species to trigger if any of the conditions are true.
Examples
TRIGGER:A:LOGICQUALIFICAT ION OR species to trigger if any of the
conditions a
re true.
2-876 MSO54, MSO56, MSO58, MSO58LP Programmer