Commands listed in alphabetical order
DISplay:SELe
ct:VIEW
This command sets or queries the selected v iew.
Group
Display Control
Syntax
DISplay:SELect:VIEW
{WAVEVIEW1|MATHFFT<x>|PLOT VIEW<x>|REFFFT<x>}
Arguments
Arguments are the selected view.
Examples
DISplay:SELect:VIEW PLOTVIEW1 selects PLOTVIEW1 as the selected view.
DISplay:SELect:VIEW? might return :DISPLAY:SELECT:VIEW MATHFFT 1
indicating MATHFFT 1 is the selected view.
DISplay:SELect:WAVEView<x>:SOUrce
This command sets or queries the selected source in the given waveview.
Group
Display Control
Syntax
DISplay:SELect:WAVEView<x> :SOUrce
{CH<x>|MATH<x>|BUS<x>|REF< x>|PLOT<x>}
Arguments
Arguments are the selected source.
Examples
DISplay:SELect:WAVEView1:S OUrce MATH1 sets MATH1 as the selected
source.
DISplay:SELect:WAVEView1:S OUrce? might return
:DISPLAY:SELECT:WAVEVIEW1: SOURCE CH2 indicating CH2 is the
selected source.
DISplay:VARpersist
This command sets or queries display persistence decay time, which is the
approximate decay time for a freshly struck persistence sample.
Group
Display Control
MSO54, MSO56, MSO58, MSO58LP Programmer 2-281