Commands listed in alphabetical order
Group
Measurement
Syntax
MEASUrement:MEAS<x>:REFLev els<x>:ABSolute:HYSTeresis <NR3>
MEASUrement:MEAS<x>:REFLev els<x>:ABSolute:HYSTeresis?
Arguments
<NR3> is the hysteresis value used for autoset.
Examples
MEASUrement:MEAS1:REFLevel s1:ABSolute:HYSTeresis 25 .0E-3 sets
the hysteresis to 25 mV.
MEASUrement:MEAS1:REFLevel s1:ABSolute:HYSTeresis? might return
:MEASUREMENT:MEAS1:REFLEVE LS1:ABSOLUTE:HYSTERESIS 3 0.0000E-3
indicating the hysteresis is set to 30 mV.
MEASUrement:MEAS<x>:REFLevels<x>:ABSolute:RISEHigh
This command sets or queries the value used as the high reference level of the
rising edge when the measurement's ref level method is set to absolute. The
measurement number is specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:REFLev els<x>:ABSolute:RISEHigh <NR3>
MEASUrement:MEAS<x>:REFLev els<x>:ABSolute:RISEHigh?
Arguments
<NR3> is the high reference level, and is the zero percent level when the
measurement's Ref level method is set to Absolute.
Examples
MEASUrement:MEAS1:REFLevel s1:ABSolute:RISEHigh 1.50 sets the
high reference level to 1.5 V.
MEASUrement:MEAS1:REFLevel s1:ABSolute:RISEHigh? might return
:MEASUREMENT:MEAS1:REFLEVE LS1:ABSOLUTE:RISEHIGH 1.0 000
indicating the high re ference level is set to 1.0 V.
MEASUrement:MEAS<x>:REFLevels<x>:ABSolute:RISELow
This command sets or que ries the value used as the low reference l evel of the
rising edge when the measurement's ref level method is set to absolute. The
measurement number is specified by x.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-471