Commands listed in alphabetical order
MEASUrement:
REFLevels:ABSolute:RISELow
This command sets or que ries the value used as the low reference l evel of the
rising edge when the measurement's ref level method is set to absolute.
Group
Measurement
Syntax
MEASUrement:REFLevels:ABSo lute:RISELow <NR3>
MEASUrement:REFLevels:ABSo lute:RISELow?
Arguments
<NR3> is the value used as the the low reference level of the rising edge
Examples
MEASUrement:REFLevels:ABSo lute:RISELow -1.5 sets the low reference
level to -1.5 V.
MEASUrement:REFLevels:ABSo lute:RISELow? might return
:MEASUREMENT:REFLEVELS:ABS OLUTE:RISELOW -1.0000 indicating the
low reference level is -1.0 V.
MEASUrement:REFLevels:ABSolute:RISEMid
This command sets or queries the value used as the mid reference level of the
rising edge when the measurement's ref level method is set to absolute.
Group
Measurement
Syntax
MEASUrement:REFLev
els:ABSolute:RISEMid
<NR3>
MEASUrement:REFLevels:ABSo lute:RISEMid?
Arguments
<NR3> is the mid reference level of the rising edge.
Examples
MEASUrement:REFLevels:ABSo lute:RISEMid 10.0E-3 sets the mid
reference to 10.0 mV.
MEASUrement:REFLevels:ABSo lute:RISEMid? might return
:MEASUREMENT:REFLEVELS:ABS OLUTE:RISEMID 0.0E+0 indicating the mid
reference is 0.0 V.
MEASUrement:REFLevels:ABSolute:TYPE
This command sets or queries the reference level t ype for the measurement.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-499