Commands listed in alphabetical order
TRIGGER:A:LOG
ICQUALIFICATION?
might return
:TRIGGER:A:LOGICQUALIFICAT ION AND indicating the instrument will
trigger if all conditions are true.
TRIGger:A:M
ODe
This command sets or queries the A trigger mode. This command is equivalent to
pushing the Mode button on the front panel.
Group
Trigger
Syntax
TRIGger:A:MODe {AUTO|NORMal}
TRIGger:A:MODe?
Arguments
AUTO generates a trigger if one is not detected within a specified time period.
NORMal waits for a valid trigger event.
Examples
TRIGGER:A:MODE NORMAL sets the trigger mode to normal.
TRIGGER:A:MODE? might return :TRIGGER:A:MODE AUTO, indicating that
the trigger mode is auto.
TRIGger:AUXLevel
For those instruments that have an Auxiliary Input (such as an MSO58LP), this
command sets or queries the Auxiliary Input voltage level to use for an edge
trigger.
Group
Trigger
Syntax
TRIGger:AUXLevel {<NR3>|ECL|TTL}
TRIGger:AUXLevel?
Arguments
<NR3> is trigger level in Volts.
ECL sets trigger level to -1.3 Volts.
TTL sets trigger level to 1.4 Volts.
Returns
<NR3> is trigger level in Volts.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-877