2.9 Controller Failure #4—Controller Cooling Fan Failure
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 65
2
2.9 Controller Failure #4—Controller Cooling Fan Failure
1. Access the Data Summary (1 of 2) screen (see “Diagnostic Menus”
on page 2-35). The Control Fan Speed meets the specification of
2500 rpm ± 50%.
Yes No
↓→Replace the controller assembly’s fan assembly (refer to
procedure 4.16 on page 4-45). If this solves the problem,
go to “Final Actions” on page 2-24. Otherwise, go to step 2.
2. Replace the interface P.C. board assembly (refer to procedure 4.18
on page 4-49). This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
3. Go to “Final Actions” on page 2-24.