2.46 Remove Skin 2 Probe Alarm Sounds
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 115
2
2.46 Remove Skin 2 Probe Alarm Sounds
1. Perform the following:
a. Disconnect the skin probe from the 2 connector on the sensor
module assembly.
b. Check the unit’s mode of operation.
The unit is in Air Mode.
Yes No
↓→Place the unit in Air Mode. If this solves the problem, go to
“Final Actions” on page 2-24. Otherwise, call Technical
Support at (800) 437-2437 for assistance.
2. Press the Silence/Reset key, and then enter Skin Mode. This
solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
3. Go to “Final Actions” on page 2-24.