2.16 Controller Failure #13—Heater Circuit Failure Detected
Chapter 2: Troubleshooting Procedures
Page 2 - 72 Isolette® Infant Incubator (Model C2000) Service Manual
2.16 Controller Failure #13—Heater Circuit Failure Detected
1. Replace the interface P.C. board assembly (refer to procedure 4.18
on page 4-49). This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
2. Go to “Final Actions” on page 2-24.