4.29 Heat Shield Latch
Chapter 4: Removal, Replacement, and Adjustment Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 4 - 81
4
Figure 4-39. Heat Shield Latch
4. Remove the Nylok®
1
screw (G) that secures the heat shield latch (F)
to the access panel (C). Discard the Nylok® screw (G).
5. Remove the heat shield latch (F) from the access panel (C).
Replacement
CAUTION:
Always replace Nylok® screws; do not reuse them. Equipment damage
could occur.
1. Replace the Nylok® screw (G), and perform the removal procedure
in reverse order.
2. To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.