2.35 Air Set Temperature Cannot Be Maintained
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 101
2
2.35 Air Set Temperature Cannot Be Maintained
1. The unit is in Air Mode.
Yes No
↓→Change the mode of operation to Air Mode. If this solves
the problem, go to “Final Actions” on page 2-24.
Otherwise, go to step 2.
2. The air set temperature is set to at least 95 °F (35 °C).
Yes No
↓→Set the air set temperature to at least 95 °F (35 °C). If this
solves the problem, go to “Final Actions” on page 2-24.
Otherwise, go to step 3.
3. Go to RAP 2.18 on page 2-74. This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
4. Go to “Final Actions” on page 2-24.