Cleaning
Chapter 6: General Procedures
Page 6 - 6 Isolette® Infant Incubator (Model C2000) Service Manual
Figure 6-1. Releasing the Inner Wall
Using an appropriate cleanser/disinfectant, thoroughly clean all
surfaces of the hood, including the sensor module, the heat shields (A),
and the access doors (D). Clean all holes and indentations, and then
dry with a clean cloth or paper towel.
Heater Radiator and Fan Impeller
WARNING:
Failure to clean the heater radiator and fan impeller could result in
sufficient lint build-up to reduce airflow, which will affect temperature
control and cause high oxygen concentrations. Infant injury could occur.
Remove any lint build-up on the heater radiator and fan impeller.
Mattress Tray, Main Deck, Heater/Impeller Cover, and Mattress Tilt
Bars
Using an appropriate cleanser/disinfectant, thoroughly clean all
surfaces, and then dry with a clean cloth or paper towel.