7.2 Oxygen System
Chapter 7: Accessories
Page 7 - 14 Isolette® Infant Incubator (Model C2000) Service Manuall
4. Install the sensor module assembly (A) (see figure 7-3 on page 7-8)
(refer to procedure 4.2 on page 4-6).
5. Activate the oxygen system, and select the 100% calibration level.
6. Calibrate the oxygen system at 100% (refer to procedure 6.2 on
page 6-20).
7. To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.
Removal
Perform the installation procedure in reverse order.