2.44 Poor Air Circulation Exists Within the Incubator
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 113
2
2.44 Poor Air Circulation Exists Within the Incubator
1. Check the air vents around the interior of the hood assembly for
obstruction. The air vents are free from obstruction.
Yes No
↓→Remove the obstruction from the air vents. If this solves
the problem, go to “Final Actions” on page 2-24.
Otherwise, go to step 2.
2. Replace the fan motor (refer to procedure 4.9 on page 4-24). This
solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
3. Go to “Final Actions” on page 2-24.