2.2 Variable Height Adjustable (VHA) Pedestal/Stand Does Not Move Up or Down
Chapter 2: Troubleshooting Procedures
Page 2 - 52 Isolette® Infant Incubator (Model C2000) Service Manual
2.2 Variable Height Adjustable (VHA) Pedestal/Stand Does Not
Move Up or Down
1. Check the variable height adjustable (VHA) pedestal stand:
a. Turn on the main Power switch (A) (see figure 2-18 on page
2-53).
b. Use your foot to press the Up arrow of the up/down switch (B) to
raise the VHA pedestal/stand to its maximum height.
c. Press and hold the Down arrow of the up/down switch (B) to
lower the VHA pedestal/stand to its minimum height.
d. Repeat using the rear up/down switch (B).
The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
2. Inspect the main Power switch (A). The main power switch (A) is in
the On position.
Yes No
↓→Turn on the main Power switch (A). If this solves the
problem, go to “Final Actions” on page 2-24. Otherwise, go
to step 3.
3. Inspect the power cord (C). The unit is plugged into an appropriate
power source.
Yes No
↓→Plug the unit into an appropriate power source. If this
solves the problem, go to “Final Actions” on page 2-24.
Otherwise, call Technical Support at (800) 437-2437 for
assistance.
4. Call Technical Support at (800) 437-2437 for assistance.