2.19 Humidity Heater Draws Too Much Current
Chapter 2: Troubleshooting Procedures
Page 2 - 76 Isolette® Infant Incubator (Model C2000) Service Manual
2.19 Humidity Heater Draws Too Much Current
1. Replace the evaporator reservoir assembly (refer to procedure 7.3
on page 7-15). This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
2. Go to “Final Actions” on page 2-24.