Preventive Maintenance
Chapter 6: General Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 6 - 11
6
Preventive Maintenance Schedule
Table 6-2. Preventive Maintenance Schedule
Function Procedure
Overall perfor-
mance
To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.
Patient compart-
ment
Inspect the patient compartment for signs of breakage.
Replace assemblies as necessary before placing the incubator
into service.
Oxygen sensor
cells
Replace both oxygen sensor cells annually (refer to procedure
4.3 on page 4-9).
Current leakage Perform the leakage current test (refer to procedure 6.5 on
page 6-27)