2.4 Oxygen Concentration is Low
Chapter 2: Troubleshooting Procedures
Page 2 - 56 Isolette® Infant Incubator (Model C2000) Service Manual
2.4 Oxygen Concentration is Low
1. Perform the following:
a. If the Low Oxygen % alarm sounds, press the Silence/Reset
softkey (A) to silence it for 4 minutes (see figure 2-20 on page
2-56).
Figure 2-20. Hood/Shell Assembly
b. Inspect the oxygen source.
The oxygen source is securely connected to the oxygen input
connector (B).
Yes No
↓→Securely connect the oxygen source to the oxygen input
connector (B). If this solves the problem, go to “Final
Actions” on page 2-24. Otherwise, go to step 2.
2. Check all access doors (C) and iris entry ports (D). All access doors
(C) and iris entry ports (D) are closed.
Yes No
↓→Close all access doors (C) and iris entry ports (D). If this
solves the problem, go to “Final Actions” on page 2-24.
Otherwise, go to step 3.