6.6 Disassembly and Assembly for Cleaning
Chapter 6: General Procedures
Page 6 - 32 Isolette® Infant Incubator (Model C2000) Service Manual
18.Remove the air intake microfilter cover and the air intake microfilter
from the unit (refer to procedure 4.19 on page 4-51).
Cleaning
For details on component cleaning, refer to “Cleaning” on page 6-3.
Assembly
NOTE:
Before assembling cleaned components into the incubator, inspect
them for any breakage or cracks. Harsh cleaning agents may harm
some of the plastics used in the patient compartment.
1. Install the impeller assembly (refer to procedure 4.8 on page 4-22).
2. Install the heater assembly and the heater radiator (refer to
procedure 4.20 on page 4-52)
3. Install the heater/impeller cover (H) (see figure 6-8 on page 6-30).
4. Install the main deck (G).
5. Slide the mattress tilt bars (F) through their openings in the main
deck (G).
6. Install the mattress tray (D) and the x-ray tray (C) (refer to procedure
4.7 on page 4-20).
7. If the unit is equipped with a weighing scale, install it (refer to
procedure 7.1 on page 7-5).
8. Place a new disposable mattress cover over the mattress (E).
9. Install the mattress (E) on the mattress tray (D).
10.Install disposable iris entry port sleeves (refer to procedure 4.11 on
page 4-33).
11. If the unit is equipped with a humidity system, perform the following:
a. Install the reservoir valve (J), the spring (K), and the seal (L) in
the humidity reservoir (I) (see figure 6-9 on page 6-31).
b. Install the lid (M) on the humidity reservoir (I).
c. Install the humidity reservoir (I) on the humidity tray (N).