EasyManuals Logo

Dräger Isolette C2000 Service Manual

Dräger Isolette C2000
524 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
Safety Tips
Chapter 1: Introduction
Isolette® Infant Incubator (Model C2000) Service Manual Page 1 - 19
1
WARNING:
Make sure that the oxygen supply to the incubator is turned off and that
the incubator is disconnected from the oxygen supply when performing
cleaning procedures. A fire and explosion hazard exists when cleaning
in an oxygen-enriched environment.
WARNING:
Failure to clean the heater radiator and fan impeller could result in
sufficient lint build-up to reduce airflow, which will affect temperature
control and cause high oxygen concentrations. Infant injury could occur.
WARNING:
Only facility-authorized personnel should perform preventive
maintenance on the Model C2000 Isolette® Infant Incubator.
Preventive maintenance performed by unauthorized personnel could
result in personal injury or equipment damage.
WARNING:
Make sure that the oxygen supply to the incubator is turned off and that
the incubator is disconnected from the oxygen supply when performing
maintenance procedures. A fire and explosion hazard exists when
performing maintenance procedures in an oxygen-enriched
environment.
WARNING:
The hood/shell assembly must attach to the pedestal/stand using the
screws provided. Failure to do so could result in the hood/shell
assembly separating from the pedestal stand when sufficiently tilted,
particularly with the hood open. Personal injury or equipment damage
could occur.
WARNING:
To keep the incubator from sliding when parked on an incline, the
pedestal stand’s front locking casters must be facing down the incline
and locked.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Isolette C2000 and is the answer not in the manual?

Dräger Isolette C2000 Specifications

General IconGeneral
BrandDräger
ModelIsolette C2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals