Safety Tips
Chapter 1: Introduction
Isolette® Infant Incubator (Model C2000) Service Manual Page 1 - 19
1
WARNING:
Make sure that the oxygen supply to the incubator is turned off and that
the incubator is disconnected from the oxygen supply when performing
cleaning procedures. A fire and explosion hazard exists when cleaning
in an oxygen-enriched environment.
WARNING:
Failure to clean the heater radiator and fan impeller could result in
sufficient lint build-up to reduce airflow, which will affect temperature
control and cause high oxygen concentrations. Infant injury could occur.
WARNING:
Only facility-authorized personnel should perform preventive
maintenance on the Model C2000 Isolette® Infant Incubator.
Preventive maintenance performed by unauthorized personnel could
result in personal injury or equipment damage.
WARNING:
Make sure that the oxygen supply to the incubator is turned off and that
the incubator is disconnected from the oxygen supply when performing
maintenance procedures. A fire and explosion hazard exists when
performing maintenance procedures in an oxygen-enriched
environment.
WARNING:
The hood/shell assembly must attach to the pedestal/stand using the
screws provided. Failure to do so could result in the hood/shell
assembly separating from the pedestal stand when sufficiently tilted,
particularly with the hood open. Personal injury or equipment damage
could occur.
WARNING:
To keep the incubator from sliding when parked on an incline, the
pedestal stand’s front locking casters must be facing down the incline
and locked.