2.33 Sensor Module Lock Does Not Secure the Sensor Module (Units Equipped With the Sensor
Module Lock Only)
Chapter 2: Troubleshooting Procedures
Page 2 - 98 Isolette® Infant Incubator (Model C2000) Service Manual
2.33 Sensor Module Lock Does Not Secure the Sensor Module
(Units Equipped With the Sensor Module Lock Only)
1. Check the area around the sensor module (A) for obstruction (see
figure 2-34 on page 2-98). The sensor module (A) is free from
obstruction.
Figure 2-34. Sensor Module
Yes No
↓→Remove the obstruction from the sensor module (A). If this
solves the problem, go to “Final Actions” on page 2-24.
Otherwise, go to step 2.
2. Replace the sensor module lock (B) (refer to procedure 4.33 on
page 4-94). This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
3. Go to “Final Actions” on page 2-24.