2.34 Incubator Takes Longer Than One Hour to Warm Up to the Air Set Temperature
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 99
2
2.34 Incubator Takes Longer Than One Hour to Warm Up to the Air
Set Temperature
1. Close all access panels (A) and access doors (B) (see figure 2-35
on page 2-99). This solves the problem.
Figure 2-35. Hood Assembly
Yes No
↓→Go to step 3.
2. Go to “Final Actions” on page 2-24.
3. Check all access grommets (C). All access grommets (C) are
securely installed.
Yes No
↓→Securely install access grommets (C) (refer to procedure
4.34 on page 4-96). If this solves the problem, go to “Final
Actions” on page 2-24. Otherwise, go to step 4.
4. All access grommets (C) are in good condition.