4.12 Access Door Latch
Chapter 4: Removal, Replacement, and Adjustment Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 4 - 37
4
Replacement
CAUTION:
Always replace Nylok®
1
screws; do not reuse them. Equipment
damage could occur.
1. Install two new Nylok® screws (G) to secure the access door latch
mount (E) and the access door latch (F) to the access panel (C).
2. Close and latch the access panel (C).
3. To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.
1. Nylok® is a registered trademark of Nylok Fastener Corporation.