6.4 Weighing Scale Calibration
Chapter 6: General Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 6 - 25
6
6.4 Weighing Scale Calibration
Tools required: Weight, 11 lb (5 kg) (P/N 03 310 16)
Using a calibrated weigh, perform the following procedure to calibrate
the scale upon installation and every 6 months thereafter:
1. Ensure that the mattress tray (A) is level and not in the
Trendelenburg or Reverse Trendelenburg position (see figure 6-6
on page 6-25).
Figure 6-6. Scale Calibration
2. Select Display 2, and press the Weigh softkey.
3. If objects are on the mattress (B), remove them.
4. At Display 2, perform the following:
a. Press the Zero softkey two times.
b. Press the Cal softkey.
5. Wait for a weight reading and the 5 kg symbol to appear.