EasyManuals Logo

Dräger Isolette C2000 Service Manual

Dräger Isolette C2000
524 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #464 background imageLoading...
Page #464 background image
6.4 Weighing Scale Calibration
Chapter 6: General Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 6 - 25
6
6.4 Weighing Scale Calibration
Tools required: Weight, 11 lb (5 kg) (P/N 03 310 16)
Using a calibrated weigh, perform the following procedure to calibrate
the scale upon installation and every 6 months thereafter:
1. Ensure that the mattress tray (A) is level and not in the
Trendelenburg or Reverse Trendelenburg position (see figure 6-6
on page 6-25).
Figure 6-6. Scale Calibration
2. Select Display 2, and press the Weigh softkey.
3. If objects are on the mattress (B), remove them.
4. At Display 2, perform the following:
a. Press the Zero softkey two times.
b. Press the Cal softkey.
5. Wait for a weight reading and the 5 kg symbol to appear.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Isolette C2000 and is the answer not in the manual?

Dräger Isolette C2000 Specifications

General IconGeneral
BrandDräger
ModelIsolette C2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals