6.4 Weighing Scale Calibration
Chapter 6: General Procedures
Page 6 - 26 Isolette® Infant Incubator (Model C2000) Service Manual
NOTE:
If a calibrated 11 lb (5 kg) weight is not placed on the mattress within 12
seconds after the 5 kg symbol appears, the Zeroing Failed system
prompt message appears.
6. Within 12 seconds after the weight reading, place a calibrated 11 lb
(5 kg) weight (C) on the mattress (B).
7. Wait for a weight reading. Perform one of the following:
• If a reading of 5.000 kg appears, go to step 8.
• If a reading of 5.000 kg does not appear, go to RAP 2.47 on page 2-
114.
8. Remove the weight (C).
9. Press the Home softkey.
10.To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.