EasyManua.ls Logo

Dräger Isolette C2000 - Page 465

Dräger Isolette C2000
524 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6.4 Weighing Scale Calibration
Chapter 6: General Procedures
Page 6 - 26 Isolette® Infant Incubator (Model C2000) Service Manual
NOTE:
If a calibrated 11 lb (5 kg) weight is not placed on the mattress within 12
seconds after the 5 kg symbol appears, the Zeroing Failed system
prompt message appears.
6. Within 12 seconds after the weight reading, place a calibrated 11 lb
(5 kg) weight (C) on the mattress (B).
7. Wait for a weight reading. Perform one of the following:
If a reading of 5.000 kg appears, go to step 8.
If a reading of 5.000 kg does not appear, go to RAP 2.47 on page 2-
114.
8. Remove the weight (C).
9. Press the Home softkey.
10.To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.

Table of Contents

Related product manuals