EasyManua.ls Logo

Dräger Isolette C2000 - 100% Calibration Fixture Installation

Dräger Isolette C2000
524 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.2 Oxygen System
Chapter 7: Accessories
Isolette® Infant Incubator (Model C2000) Service Manual Page 7 - 13
7
16.Calibrate the oxygen system at 21% (refer to procedure 6.2 on page
6-20).
17.To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.
100% Calibration Fixture Installation
1. Remove the sensor module assembly (A) from the unit (see figure
7-3 on page 7-8) (refer to procedure 4.2 on page 4-6).
NOTE:
The 21% fixture is located under the sensor module opening.
2. Remove the two screws (U) that secure the 21% calibration fixture
to the hood assembly (V) (see figure 7-7 on page 7-13). Retain the
two screws (U).
Figure 7-7. 100% Oxygen Calibration Fixture
3. Install the two screws (U) to secure the slide lock (W) and the 100%
calibration fixture (X) directly under the opening in the hood
assembly (V).

Table of Contents

Related product manuals