2.18 Heater Thermocouple Does Not Work Properly
Chapter 2: Troubleshooting Procedures
Page 2 - 74 Isolette® Infant Incubator (Model C2000) Service Manual
2.18 Heater Thermocouple Does Not Work Properly
1. Perform the following:
a. Remove the two wingnuts (A) and the two flat washers (B) that
secure the controller assembly (C) in the unit (see figure 2-24 on
page 2-74).
Figure 2-24. Controller Assembly
b. Slide the controller assembly (C) from the unit (refer to
procedure 4.4 on page 4-11).
c. Check the connection of the heater assembly’s wiring harness
assembly (D) at the interface/power supply module in the
controller assembly (C).
The heater assembly’s wiring harness assembly (D) is securely
connected at the controller assembly (C).
Yes No
↓→Securely connect the heater assembly’s wiring harness
assembly (D) at the interface/power supply module in the
controller assembly (C). If this solves the problem, go to
“Final Actions” on page 2-24. Otherwise, go to step 2.