4.7 Mattress, Mattress Tray, and X-ray Tray
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 20 Isolette® Infant Incubator (Model C2000) Service Manual
4.7 Mattress, Mattress Tray, and X-ray Tray
Tools required: None
Removal
1. Lift the hood assembly, and carefully tilt it back to its open position.
2. Remove the mattress (A) from the mattress tray (B) (see figure 4-7
on page 4-20).
Figure 4-7. Mattress Platform Assembly
3. If necessary, remove the mattress tray (B) and the x-ray tray (C)
from the unit.
4. If necessary, remove the x-ray tray (C) from the mattress tray (B).