2.3 Skin Set Temperature Cannot Be Achieved or Maintained
Chapter 2: Troubleshooting Procedures
Page 2 - 54 Isolette® Infant Incubator (Model C2000) Service Manual
2.3 Skin Set Temperature Cannot Be Achieved or Maintained
1. The Low Skin Temperature alarm sounds.
Yes No
↓→Go to RAP 2.37 on page 2-103.
2. Perform the following:
a. Silence the Low Skin Temperature alarm for 15 minutes by
pressing the Silence/Reset softkey (A) (see figure 2-19 on page
2-54).
Figure 2-19. Hood/Shell Assembly
b. Check all access doors (B) and iris entry ports (C).
All access doors (B) and iris entry ports (C) are closed.
Yes No
↓→Close all access doors (B) and iris entry ports (C). If this
solves the problem, go to “Final Actions” on page 2-24.
Otherwise, go to step 3.
3. Check the unit’s mode of operation. The unit is in Skin Mode.