2.14 Controller Failure #10—Watchdog Timer Failure
Chapter 2: Troubleshooting Procedures
Page 2 - 70 Isolette® Infant Incubator (Model C2000) Service Manual
2.14 Controller Failure #10—Watchdog Timer Failure
1. Replace the electroluminescent (EL) central processing unit (CPU)
P.C. board assembly (refer to procedure 4.15 on page 4-43). This
solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
2. Go to “Final Actions” on page 2-24.