4.13 Access Door and Access Door Pivot Hinges
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 40 Isolette® Infant Incubator (Model C2000) Service Manual
CAUTION:
Always replace Nylok®
1
screws; do not reuse them. Equipment
damage could occur.
3. Install the four new Nylok® screws (F) to secure the access door
pivot hinges (G) to the access panel (D).
4. Close and latch the access panel (D).
5. To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.
1. Nylok® is a registered trademark of Nylok Fastener Corporation.