2.37 Low Skin Temperature Alarm or High Skin Temperature Alarm Does Not Sound When Skin
Temperature Drops or Rises Out of Specification
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 103
2
2.37 Low Skin Temperature Alarm or High Skin Temperature Alarm
Does Not Sound When Skin Temperature Drops or Rises Out
of Specification
1. The unit is in Skin Mode.
Yes No
↓→Change the mode of operation to Skin Mode. If this solves
the problem, go to “Final Actions” on page 2-24.
Otherwise, go to step 2.
2. Replace the interface P.C. board assembly (refer to procedure 4.18
on page 4-49). The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
3. Replace the electroluminescent (EL) central processing unit (CPU)
P.C. board (refer to procedure 4.15 on page 4-43). This solves the
problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
4. Go to “Final Actions” on page 2-24.