2.17 Controller Failure #14—Low Heater Current Detected
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 73
2
2.17 Controller Failure #14—Low Heater Current Detected
1. Access the Data Summary (2 of 2) screen (see “Diagnostic Menus”
on page 2-35). The AC Current meets the specification of 3 A.
Yes No
↓→Replace the heater assembly (refer to procedure 4.20 on
page 4-52). If this solves the problem, go to “Final Actions”
on page 2-24. Otherwise, go to step 2.
2. Replace the interface P.C. board assembly (refer to procedure 4.18
on page 4-49). This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
3. Go to “Final Actions” on page 2-24.