Final Actions
Chapter 2: Troubleshooting Procedures
Page 2 - 24 Isolette® Infant Incubator (Model C2000) Service Manual
b. Place the probe of a calibrated hygrometer inside the hood
assembly at 4" (10 cm) above the center mattress.
c. If necessary, select Air Mode.
d. Pre-warm the incubator to 95 F (35 C).
e. At Display 1, press the Humidity softkey to turn on the humidity.
f. Set the humidity setpoint to 50%.
NOTE:
The humidity system sensor has a tolerance of ± 6% relative humidity
(RH). Some hygrometers have an additional tolerance of ± 3% RH.
When using a hygrometer to determine the performance of the humidity
system, allow for a total of ± 6% RH, plus the hygrometer tolerance.
Within 30 to 60 minutes, the hygrometer and Humidity display both
read 50% ± 6%.
Yes No
↓→Go to RAP 2.40 on page 2-106.
45.Go to “Final Actions” on page 2-24.
Final Actions
1. Complete the required preventive maintenance procedures. See
“Preventive Maintenance Checklist” on page 6-12.
2. Complete all required administrative tasks.