2.31 Air Intake Is Not Working Properly
Chapter 2: Troubleshooting Procedures
Page 2 - 94 Isolette® Infant Incubator (Model C2000) Service Manual
2.31 Air Intake Is Not Working Properly
1. Perform the following:
a. Loosen the two thumbscrews (A) on the air intake filter cover (B),
and remove the air intake filter cover (B) (see figure 2-32 on
page 2-94).
Figure 2-32. Air Intake Microfilter
b. Inspect the air intake filter cover (B).
The air intake filter cover (B) is free from obstruction.
Yes No
↓→Remove the obstruction from the air intake filter cover (B).
If this solves the problem, go to “Final Actions” on page
2-24. Otherwise, go to step 2.