2.30 Mattress Tray Is Damaged
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 93
2
Yes No
↓→Replace the mattress tray (A) (refer to procedure 4.7 on
page 4-20). If this solves the problem, go to “Final Actions”
on page 2-24. Otherwise, call Technical Support at (800)
437-2437 for assistance.
2. Perform the following:
a. Return the mattress tray (A) to its original position.
b. Close the access panel, and secure it using its blue slide access
panel latches or its access panel pawl latch knobs.
c. Go to “Final Actions” on page 2-24.