2.38 When Skin Probes are Disconnected, Connect Skin 1 Probe Alarm Does Not Sound
Chapter 2: Troubleshooting Procedures
Page 2 - 104 Isolette® Infant Incubator (Model C2000) Service Manual
2.38 When Skin Probes are Disconnected, Connect Skin 1 Probe
Alarm Does Not Sound
1. The unit is in Skin Mode.
Yes No
↓→Change the mode of operation to Skin Mode. If this solves
the problem, go to “Final Actions” on page 2-24.
Otherwise, go to step 2.
2. Go to RAP 2.42 on page 2-109.