2.47 Scale Is Not Working Properly
Chapter 2: Troubleshooting Procedures
Page 2 - 116 Isolette® Infant Incubator (Model C2000) Service Manual
2.47 Scale Is Not Working Properly
1. Inspect the mattress tray (A) (see figure 2-38 on page 2-117). The
mattress tray (A) is level.
Yes No
↓→Using the mattress tilt mechanism knobs (B), place the
mattress tray (A) in the level position. If this solves the
problem, go to “Final Actions” on page 2-24. Otherwise, go
to step 2.
2. The Too Much Weight alarm sounds.
Yes No
↓→Go to step 6.
3. Perform the following:
a. Silence the Too Much Weight alarm for 5 minutes by pressing
the Silence/Reset key (C).
b. Remove the excess weight from the mattress (D).
The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
4. The Scale Disconnected alarm sounds.
Yes No
↓→Go to step 6.
5. Go to step 10.
6. Perform the following:
a. If you desire a pounds/ounces display, refer to “Installation and
Set-up” on page 6-13.
b. Remove any objects from the mattress (D).
c. Select Display 2, and press the Weigh softkey.
d. Press the Zero softkey twice in succession.
The Weight display reads 0, and the Weight Sample bar begins to
fill.
Yes No
↓→Remove all objects from the mattress (D). If this solves the
problem, go to “Final Actions” on page 2-24. Otherwise, go
to step 10.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Isolette C2000 and is the answer not in the manual?

Dräger Isolette C2000 Specifications

General IconGeneral
BrandDräger
ModelIsolette C2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals