4.1 Skin Temperature Probe
Chapter 4: Removal, Replacement, and Adjustment Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 4 - 5
4
4.1 Skin Temperature Probe
Tools required: None
Removal
Disconnect the skin temperature probe (A) from the sensor module
assembly (B) (see figure 4-1 on page 4-5).
Figure 4-1. Skin Temperature Probe
Replacement
1. Perform the removal procedure in reverse order.
2. To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.