Function Checks
Chapter 2: Troubleshooting Procedures
Page 2 - 12 Isolette® Infant Incubator (Model C2000) Service Manual
d. Close each access panel (I), and secure them using the blue
slide access panel latches (J) or the access panel pawl latch
knobs (K).
The blue slide access panel latches (J) or the access panel pawl
latch knobs (K) are properly secured to avoid accidental opening of
the access panel(s) (I).
Yes No
↓→Go to RAP 2.25 on page 2-83.
12.The hood assembly (D) is equipped with an iris entry port (L) (see
figure 2-3 on page 2-12).
Yes No
↓→Go to step 14.
Figure 2-3. Iris Entry Port and Access Doors
13.Check the iris entry port (L) by rotating its outer ring (N). The iris
entry port (L) opens and closes as rotation is continued through
360 .
Yes No
↓→Go to RAP 2.26 on page 2-85.
14.The hood assembly (D) is equipped with an access door (O).