Function Checks
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 13
2
Yes No
↓→Go to step 17.
15.Press the door release (Q) of each access door (O). Each access
door (O) swings open.
Yes No
↓→Go to RAP 2.27 on page 2-86.
16.Close the access doors (O), and check for proper latching and
quietness. The gaskets (R) of each access door (O) are placed
properly on the inner and outer walls.
Yes No
↓→Go to RAP 2.27 on page 2-86.
17.Perform the following:
a. On a unit with a Series 00 hood assembly (D), slide the blue
slide access panel latches (J), and open the access panel (I)
(see figure 2-2 on page 2-11).
or
On a unit with a hood assembly (D) later than Series 00, rotate
the access panel pawl latch knobs (K), and open the access
panel (I).
b. Inspect the heat shield latches (S) and the heat shield (T) (see
figure 2-4 on page 2-14).