2.43 Oxygen System Is Not Working Properly (Units Equipped With an Oxygen System Only)
Chapter 2: Troubleshooting Procedures
Page 2 - 112 Isolette® Infant Incubator (Model C2000) Service Manual
2.43 Oxygen System Is Not Working Properly (Units Equipped With
an Oxygen System Only)
1. The Oxygen Solenoid Fail alarm sounds.
Yes No
↓→Go to step 3.
2. Replace the oxygen solenoid (refer to procedure 7.2 on page 7-8).
The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
3. Perform the following:
a. Press the Silence/Reset softkey to silence the Oxygen
Calibration Required alarm for 5 minutes, if necessary.
b. Calibrate the oxygen sensors (refer to procedure 6.2 on page
6-20).
The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
4. Replace the oxygen sensors (refer to procedure 7.2 on page 7-8).
The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
5. For assistance, call Technical Support at (800) 437-2437.