4.11 Iris Entry Port Sleeve
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 34 Isolette® Infant Incubator (Model C2000) Service Manual
4. Rotate the outer ring (F) in the other direction:
If the iris entry port sleeve (A) is properly installed, it opens with the
rotation of the outer ring (F).
If the iris entry port sleeve (A) does not open with the rotation of the
outer ring (F), remove and re-install it (see “Removal” on page
4-33).
5. To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.