2.27 Access Door Does Not Operate or Latch Properly
Chapter 2: Troubleshooting Procedures
Page 2 - 86 Isolette® Infant Incubator (Model C2000) Service Manual
2.27 Access Door Does Not Operate or Latch Properly
1. Inspect each access door (A) (see figure 2-28 on page 2-86). The
access doors (A) are free from cracks and in good condition.
Figure 2-28. Access Door
Yes No
↓→Replace the access door (A) (refer to procedure 4.13 on
page 4-38). If this solves the problem, go to “Final Actions”
on page 2-24. Otherwise, go to step 2.
2. Inspect the area around the access door (A) for possible
obstruction. The area around the access door (A) is free from
obstruction.
Yes No
↓→Remove the obstruction from the access door (A). If this
solves the problem, go to “Final Actions” on page 2-24.
Otherwise, go to step 3.
3. Close each access door (A). Each access door (A) closes properly.