2.27 Access Door Does Not Operate or Latch Properly
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 87
2
Yes No
↓→Replace the access door gasket (B) (refer to procedure
4.27 on page 4-76). If this solves the problem, go to “Final
Actions” on page 2-24. Otherwise, go to step 4.
4. Press the access door latch (C) at each access door (A). Each
access door (A) swings open.
Yes No
↓→Replace the access door latch (C) (refer to procedure 4.12
on page 4-35). If this solves the problem, go to “Final
Actions” on page 2-24. Otherwise, go to step 5.
5. The problem is solved.
Yes No
↓→Replace the pivot hinges (D) and the torsion spring on the
access door (A) (refer to procedure 4.13 on page 4-38). If
this solves the problem, go to “Final Actions” on page 2-24.
Otherwise, call Technical Support at (800) 437-2437 for
assistance.
6. Go to “Final Actions” on page 2-24.