4.7 Mattress, Mattress Tray, and X-ray Tray
Chapter 4: Removal, Replacement, and Adjustment Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 4 - 21
4
Replacement
1. Perform the removal procedure in reverse order. Ensure that the
mattress tray (B) aligns with the mattress lift bars (D).
2. To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.