4.18 Power Supply P.C. Board Assembly and Interface P.C. Board Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 50 Isolette® Infant Incubator (Model C2000) Service Manual
5. Remove the power supply P.C. board (C) and the seven nylon
standoffs (E) from the interface P.C. board assembly (D) (see
“Component Handling” on page 6-8).
Replacement
1. Perform the removal procedure in reverse order.
2. To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.