2.23 A Stuck Key Is Detected
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 81
2
2.23 A Stuck Key Is Detected
1. Perform the following:
a. Access the Key Check screen (see “Diagnostic Menus” on page
2-35).
b. Press each key in succession.
Each box on the screen illuminates when its respective key is
pressed.
Yes No
↓→Replace the electroluminescent display (refer to procedure
4.14 on page 4-41). If this solves the problem, go to “Final
Actions” on page 2-24. Otherwise, go to step 2.
2. The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
3. For assistance, call Technical Support at (800) 437-2437.