Diagnostic Menus
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 119
2
2.48 Oxygen System’s Cal Fail System Prompt Message Appears
1. Calibrate the oxygen system (refer to procedure 6.2 on page 6-20).
The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
2. Attempt to calibrate the oxygen system again (refer to procedure 6.2
on page 6-20). The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
3. For assistance, call Technical Support at (800) 437-2437.