4.31 Mattress Tilt Knob
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 84 Isolette® Infant Incubator (Model C2000) Service Manual
4.31 Mattress Tilt Knob
Tools required: Small screwdriver
Phillips head screwdriver
9/64" Allen™
1
wrench
Flat padded surface
Needle-nose pliers
Pliers
Removal
SHOCK HAZARD:
Unplug the power cord from the controller assembly. Failure to do so
could result in personal injury or equipment damage.
1. Unplug the power cord from the controller assembly (A) (see figure
4-42 on page 4-84).
Figure 4-42. Shell Assembly
1. Allen™ is a trademark of Industrial Fasteners, Inc.