2.28 Heat Shield Does Not Latch Properly
Chapter 2: Troubleshooting Procedures
Page 2 - 88 Isolette® Infant Incubator (Model C2000) Service Manual
2.28 Heat Shield Does Not Latch Properly
1. Using the heat shield latches (A), properly secure the top of the heat
shield (B) to the access panel (C) (see figure 2-29 on page 2-88).
The problem still exists.
Figure 2-29. Heat Shield
Yes No
↓→Go to “Final Actions” on page 2-24.
2. Check the area between the heat shield (B) and the access panel
(C) for possible obstruction. The area between the heat shield (B)
and access panel (C) is free from obstruction.
Yes No
↓→Remove the obstruction from between the heat shield (B)
and the access panel (C). If this solves the problem, go to
“Final Actions” on page 2-24. Otherwise, go to step 3.
3. Replace the heat shield latches (A) (refer to procedure 4.29 on page
4-80). The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.